International Journal of Engineering Technology and Management Sciences

2023, Volume 7 Issue 5

Design And Analysis Of Cardiac Stent

AUTHOR(S)

S R Rahul, T Sekar, N Nandakumar, S Periyasmay

DOI: https://doi.org/10.46647/ijetms.2023.v07i05.018

ABSTRACT
This project presents the modification and analysis of two stent designs based on an existing design. The modifications were made using SolidWorks, followed by ANSYS analysis to evaluate von Mises stress and strain, stress intensity, maximum principle stress, and total deformation. Design Expert was then utilized, employing the Response Surface Methodology (RSM) Box-Behnken method, to conclude the design.The generated ANOVA table enabled the identification of significant factors and interactions. Residual analysis was conducted to assess the model's fit and accuracy. The abstract highlights the iterative design process involving SolidWorks, ANSYS, and Design Expert, emphasizing the use of ANOVA and residual analysis to improve the stent design. The optimized stent designs exhibited improvements in stress distribution, strain, stress intensity, and deformation characteristics. This integrated approach utilizing SolidWorks, ANSYS, and Design Expert holds promise for enhancing stent performance and optimizing patient outcomes.

Page No: 154 - 161

References:

    • Akdogan, G., And O. B. Istanbullu. Damar içi stentlerde malzeme seçiminin ve tasarımının restenoz ve diger stent kaynaklı problemlere etkileri, stentlerin ekonomideki yeri (bir genel derleme). Eur. J. Sci. Technol. 204–215, 2020
    • Abraha, i.,C. Romagnoli, A. Montedori, and R. Cirocchi, “Thoracic stent graft    versus surgery for thoracic aneurysm. Cochrane database syst”. Rev. 2016.
    • Canić s., Tambaca J.”Cardiovascular stents” as PDE nets: 1D vs 3DIMA J. Appl.   Math., 77 (6) (2018), pp. 748-779.
    • Bressloff N.W., Ragkousis G., Curzen N “ Design optimisation of coronary artery stent systems  Ann”. Biomed. Eng., 44 (2) (2016),pp. 357-367
    • Doyle, B., Rihal, C. S. , O’Sullivan, C. J. , Lennon, R. J. , Wiste, H. J. , Bell, M., Bresnahan, J., and Holmes, D. R. , 2007, “Outcomes of Stent Thrombosis and Restenosis During Extended Follow-Up of Patients Treated With Bare-Metal Coronary Stents,” Circulation, 116(21), pp. 2391–2398.
    • James, S. K. , Stenestrand, U., Lindbäck, J., Carlsson, J., Scherstén, F., Nilsson, T., Wallentin, L., Lagerqvist, B., and Group, S. S. , 2009, “Long-Term Safety and Efficacy of Drug-Eluting Versus Bare-Metal Stents in Sweden,” N. Engl. J. Med., 360(19), pp. 1933–1945.Stone, G. W. , Rizvi, A., Sudhir, K., Newman, W., Applegate, R. J. , Cannon, L. A. , Maddux, J. T. , Cutlip, D. E. , Simonton, C. A. , Sood, P., Kereiakes, D. J. , and Investigators, S. I. , 2011, “Randomized Comparison of Everolimus- and Paclitaxel-Eluting Stents. 2-Year Follow-Up From the SPIRIT (Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System) IV Trial,” J. Am. Coll. Cardiol., 58(1), pp. 19–25.
    • Leon, M. B. , Kandzari, D. E. , Eisenstein, E. L. , Anstrom, K. J. , Mauri, L., Cutlip, D. E. , Nikolsky, E., O’Shaughnessy, C., Overlie, P. A. , Kirtane, A. J. , McLaurin, B. T. , Solomon, S. L. , Douglas, J. S. , Popma, J. J. , and Investigators, E. I. , 2009,“Late Safety, Efficacy, and Cost-Effectiveness of a Zotarolimus-Eluting Stent Compared With a Paclitaxel-Eluting Stent in Patients With De Novo Coronary Lesions: 2-Year Follow-Up From the ENDEAVOR IV Trial (Randomized, Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Taxus Paclitaxel-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions).” JACC Cardiovasc. Interv., 2(12), pp. 1208–1218.
    • Finn, A. V., Nakazawa, G., Joner, M., Kolodgie, F. D. , Mont, E. K. , Gold, H. K. , and Virmani, R., 2007, “Vascular Responses to Drug Eluting Stents: Importance of Delayed Healing,” Arterioscler., Thromb., Vasc. Biol., 27(7), pp. 1500–1510.
    • Joner, M., Nakazawa, G., Finn, A. V. , Quee, S. C. , Coleman, L., Acampado, E., Wilson, P. S. , Skorija, K., Cheng, Q., Xu, X., Gold, H. K. , Kolodgie, F. D. , and Virmani, R., 2008, “Endothelial Cell Recovery Between Comparator Polymer-Based Drug-Eluting Stents,” J. Am. Coll. Cardiol., 52(5), pp. 333–342.
    •  Kotani, J., Awata, M., Nanto, S., Uematsu, M., Oshima, F., Minamiguchi, H., Mintz, G. S. , and Nagata, S., 2006, C: Angioscopic Findings,” J. Am. Coll. Cardiol., 47(10), pp. 2108–2111.
    •  LaDisa, J. F. , Olson, L. E. , Molthen, R. C. , Hettrick, D. A. , Pratt, P. F. , Hardel, M. D. , Kersten, J. R. , Warltier, D. C. , and Pagel, P. S. , 2005, “Alterations in Wall Shear Stress Predict Sites of Neointimal Hyperplasia After Stent Implantation in Rabbit Iliac Arteries,” Am. J. Physiol. Heart Circ. Physiol., 288(5), pp. H2465–2475.
    •  Liu, S. Q. and Goldman, J., 2001, “Role of Blood Shear Stress in the Regulation of Vascular Smooth Muscle Cell Migration,” IEEE Trans. Biomed. Eng., 48(4), pp. 474–483.
    •  M. Agrawal et al.Classification of mechanisms of strut malapposition after angiographically optimized stent implantation: an optical coherence tomography studycatheter cardiovasc interv(2017).
    •  Sprague, E. A. , Luo, J., and Palmaz, J. C. , 2000, “Endothelial Cell Migration Onto Metal Stent Surfaces Under Static and Flow Conditions,” J. Long Term Eff. Med. Implants, 10(1–2), pp. 97–110.
    •  LaDisa, J. F. , Olson, L. E. , Guler, I., Hettrick, D. A. , Audi, S. H. , Kersten, J. R. , Warltier, D., and C.Pagel, P. S. , 2004, “Stent Design Properties and Deployment Ratio Influence Indexes of Wall Shear Stress: A Three-Dimensional Computational Fluid Dynamics Investigation Within a Normal Artery,” J. Appl. Physiol., 97(1), pp. 424–430.
    • He, Y., Duraiswamy, N., Frank, A. O. , and Moore, J. E. , 2005, “Blood Flow in Stented Arteries: A Parametric Comparison of Strut Design Patterns in three Dimensions,” ASME J. Biomech. Eng., 127(4), pp. 637–647.
    •  Duraiswamy, N., Schoephoerster, R. T. , and Moore, J. E. , 2009, “Comparison of Near-Wall Hemodynamic Parameters in Stented Artery Models,” ASME J. Biomech Eng, 131(6), p. 061006.
    • Nagendra Singh; Manoj Kumar Agrawal; Sanjeev Kumar Verma; Ashish Kumar Tiwari. "A Review on Effect of Stress and Strain Distribution on the AA5083 With Respect to Different Channel Angle of ECAP". International Research Journal on Advanced Science Hub, 4, 03, 2022, 57-66. doi: 10.47392/irjash.2022.013
    • Nagendra Singh; Manoj Kumar Agrawal; Sanjeev Kumar Verma; Ashish Kumar Tiwari. "A Review on Effect of Stress and Strain Distribution on the AA5083 With Respect to Different Channel Angle of ECAP". International Research Journal on Advanced Science Hub, 4, 03, 2022, 57-66. doi: 10.47392/irjash.2022.013
    • Sayuj G K; Rishabh Almel; Rohan G; Shashank Naik; Chandru A. "Design and Analysis of Winglets on GOE767-il Aerofoil". International Research Journal on Advanced Science Hub, 03, Special Issue 7S, 2021, 125-131. doi: 10.47392/irjash.2021.221
    • Chandan Kumar Sangewar; Chinmay Pagey; Sakshi Chauhan; Suganeshwari G. "A Review on the Movie Recommendation System Using Big Data". International Research Journal on Advanced Science Hub, 5, Issue 05S, 2023, 376-381. doi: 10.47392/irjash.2023.S051
    • Murphy, J. B. and Boyle, F. J. , 2010, “A Full-Range, Multi-Variable, CFD-Based Methodology to Identify Abnormal Near-Wall Hemodynamics in a Stented Coronary Artery,” Biorheology, 47(2), pp. 117–132.

    How to Cite This Article:
    S R Rahul, T Sekar, N Nandakumar, S Periyasmay . Design And Analysis Of Cardiac Stent . ijetms;7(5):154-161. DOI: 10.46647/ijetms.2023.v07i05.018